S P E C I F I C A T I O N S
Model Number:
6400
Tests Per Hour:
6 – 7
Operator Time Per Test:
1 Minute
Precision Classification:
0.1% Class
Jacket Type:
Isoperibol, Water Jacket
Oxygen Fill:
Automatic
Bucket Fill:
Automatic
Bomb Wash:
Automatic
Bomb Model Options:
1138, 250mL, Alloy 20
1138CL, 250mL, Alloy G30
Balance Communication:
USB
Printer Communication:
USB
Network Connection:
TCP/IP via Ethernet
Dimensions (cm):
42w x 46d x 51h
T
he 6400 Automatic Isoperibol
Calorimeter
represents the next
evolutionary step in the Parr automated
calorimeters. Inclusive and compact,
the instrument incorporates a closed
loop cooling subsystem into the calo-
rimeter. This subsystem uses a ther-
moelectric cooler assembly attached
directly to a one liter water tank which
supplies cooling water to the calorim-
eter. An external nitrogen pressurized
tank is used to supply rinse water to
the calorimeter. This model features the
fixed bomb and bucket design, allowing
for automated bucket and jacket fill as
well as automated vessel fill and rinse.
The 6400 requires one minute of opera-
tor time per test, allowing a techni-
cian to operate up to four calorimeters
simultaneously.
Quick Twist-Lock Bomb
The 1138 Oxygen Bomb has been re-
designed to withstand a higher magni-
tude of tests. The head is designed with
an O-ring groove which is optimized to
minimize frictional wear, in turn improv-
ing the lifetime of the seal. The bomb
head is removable for fast sample load-
ing using the patented Quick Twist-Lock
vessel closure design.
Laboratory Requirements
The calorimeter requires a source of
99.5% oxygen, a source of nitrogen or
house air at 80 psi, and deionized water.
6400 Automatic Isoperibol Calorimeter
6400 Automatic Isoperibol Calorimeter
P a r r I n s t r u m e n t C o m p a n y
10