6200 Isoperibol Calorimeter
S P E C I F I C A T I O N S
Model Number:
6200
Tests Per Hour:
4 – 7
Operator Time Per Test:
6 Minutes
Precision Classification:
0.05 - 0.1% Class
Jacket Type:
Isoperibol, Water Jacket
Oxygen Fill:
Automatic
Bucket Fill:
Manual
Bomb Wash:
Manual
Bomb Model Options:
1108, Alloy 20
1108CL, Alloy G30
1108B, Alloy 20
1108BCL, Alloy G30
1108BP, Alloy 20
1108BPCL, Alloy G30
1108P, Alloy 20
1108PCL, Alloy G30
1109A, 22mL Semi-micro Bomb
1104, High Strength Bomb
Balance Communication:
USB
Printer Communication:
USB
Network Connection:
TCP/IP via Ethernet
Dimensions (cm):
57w x 40d x 43h
T
he 6200 Isoperibol Calorimeter
is Parr’s most precise model. The
traditional removable bomb and bucket
design along with the water jacket of this
calorimeter afford the user with complete
control over the combustion process.
This makes the 6200 the ultimate choice
for precise research and development
and quality control work.
Isoperibol Jacket System
Outstanding thermal jacketing is provided
by a circulating water system driven by a
built-in, high capacity pump which main-
tains a continuous forced flow around the
sides and bottom of the bucket chamber
and through the cover as well. A sealed
immersion heater and a built-in heat
exchanger, both operated by the calori-
meter controller, provide precise jacket
temperature control.
Automatic Oxygen Fill
To speed and simplify the bomb filling
operation, the 6200 Calorimeter has an
automatic system for charging the bomb
with oxygen. Oxygen is connected to
a microprocessor controlled solenoid
installed in the calorimeter. To fill the
bomb, the operator simply slips the filling
hose connector onto the bomb inlet valve
and presses a button on the touch screen
to start the filling sequence.
Laboratory Requirements
The calorimeter requires a source of
99.5% oxygen and deionized water for
operation. If a closed loop system is
chosen, the water handling system must
be installed so that the fan will not blow
on the calorimeter.
Expanded System
The 6220 Expanded System includes
the following components:
• 6200 Calorimeter
• 6510 Water Handling System
• 1758 Printer
• Extra Bomb and Bucket
• 6008P Bomb Maintenance Kit
• 6209P, 1 Year Service Kit
6200 Ordering Guide
6200 Isoperibol Calorimeter
Model No.
Voltage
Description
6200EA / EF
115 V / 230 V 6200 Calorimeter with 1108P Oxygen Bomb of Alloy 20
6200CLEA / EF 115 V / 230 V 6200 Calorimeter with 1108PCL Oxygen Bomb of Alloy G30
6220EA / EF
115 V / 230 V 6220 Expanded System with two 1108P Oxygen Bombs of Alloy 20 and Principal Components
6220CLEA / EF 115 V / 230 V 6220 Expanded System with two 1108CLP Oxygen Bombs of Alloy G30 and Principal Components
6220 Expanded System
6200 Isoperibol Calorimeter
P a r r I n s t r u m e n t C o m p a n y
12